How to use Cholesterol Detector

2022-01-08

How to use Cholesterol Detector

Author: Lily    Time:2022/1/7
Baili Medical Suppliers(Xiamen) Co., is a professional medical devices supplier based in Xiamen, China. Our main products: Protective Equipment, Hospital Equipment, First Aid Equipment, Hospital and Ward Facilities.
Wash your hands and dry them. Take out the blood glucose meter. Then open the test paper box and you will find that there is a test paper bottle with the word "chip" on it. Open the test paper bottle and take out the small card, and install it on the side of the battery box on the back of the blood glucose meter. Insert the battery and close the back cover. Our Cholesterol Detector products have won the recognition of customers all over the world with their excellent quality!
1. The end with the golden conductive tape is downward. The side of the card with the chip (black rectangle) faces the battery side. After installation, the upper edge of the card is flush with the back of the blood glucose meter. Please pay attention to the installation direction of the battery, if the installation is wrong, the blood glucose meter will not work.
2. Turn on the instrument and adjust the time, measurement method and display unit according to the manual (according to the manual).
3. Take a strip of test paper from the test paper bottle and quickly close the bottle cap. Insert the test strip with the silver band up into the blood glucose meter.
4. Rotate the blood collection pen, take a disposable blood collection needle, insert the round end of the hand into the needle slot of the blood collection pen and push it firmly.
! Note: The lancet is for one-time use and cannot be reused.
5. Adjust the penetration depth of the blood sampling needle. The penetration depth varies with the thickness of the finger skin. Generally, select "2". If you feel that your blood volume is insufficient, please adjust to "3"-"5".
6. Sterilize the blood sampling position of the finger with alcohol, press the blood sampling pen on the finger after the alcohol is dry, and press the blood sampling pen button. Put down the lancet.
7. If the blood sampling depth is appropriate, there should be a drop of blood on the finger, (make sure the test paper is inserted and the blood drop is flashing on the screen of the instrument) Use the blood to touch the top of the semicircular mouth of the test paper, and the blood will be sucked into the test paper automatically.
! Note: If there is not much blood on your finger, you can press with another finger, but you can't use excessive force, otherwise the measurement result will be in error.
8. Press the blood sampling point with a dry medical cotton swab.
9. The instrument will automatically record the time after inhaling the blood, and the result will be output after 15 seconds.
Precautions:
1. Please pay attention to adjusting the penetration depth of the blood sampling needle when taking blood. If the penetration is too shallow and there is not enough bleeding, the measurement will not be possible. If you squeeze the bleeding point too hard, a large amount of tissue fluid will be contained in the blood to be measured, which will eventually lead to inaccurate measurement.
2. When dripping blood, the blood drop should be as close as possible to the top of the semicircle of the test paper, so that the blood can be sucked and measured smoothly by the test paper. If the blood cannot touch the top of the semicircle, no amount of blood will be measured.
3. When the instrument displays "low", it is mostly because the blood volume is insufficient or the blood is not sucked into the test paper.
4. The shelf life of each bottle of test paper is three months. Please cover the test paper bottle as soon as possible when taking the test paper to extend the service life of the test paper as much as possible.
5. The test paper should be placed away from light to prevent the test paper from degeneration.
6. The blood glucose meter is an electronic device and cannot be stored in the refrigerator.
7. When "Hi" appears on the screen repeatedly, it means that the blood sugar is high, please go to the hospital to see a doctor immediately.
8. In order to determine the accuracy of the results, please keep the instrument clean.
9. Do not use test strips that have expired.
10. Do not use bent, cracked or deformed test paper.
11. The unused test paper should always be kept in the original test paper bottle.
12. The test paper should be stored at 10-30 degrees Celsius, and avoid light and heat.
13. When taking the test paper, do not touch the semicircular sample application area.
14. The test paper cannot be reused.
15. The test paper taken out of the test paper bottle should be used immediately.
16. This product cannot be used for infant whole blood testing.



We use cookies to offer you a better browsing experience, analyze site traffic and personalize content. By using this site, you agree to our use of cookies. Privacy Policy